6.3.18 Low FiO
2
Alarm
Property Description
Alarm text
Low FiO2
Priority High
Alarm condition A Low FiO
2
alarm will be given when the measured FiO
2
is
below the alarm limit for 30 seconds.
Possible cause
Decreased oxygen inlet.
Disconnection of oxygen inlet.
Increased minute ventilation.
High leakage.
Setting range
21% to 100%
Off
Setting resolution 1%
Ventilator action The ventilator will continue treatment with the same settings.
6.3.19 High SpO
2
Alarm
Property Description
Alarm text
High SpO2
Priority Medium
Alarm condition A High SpO
2
alarm will be given when the measured SpO
2
exceeds the alarm limit for 30 seconds.
Possible cause Too high flow of bleed-in oxygen.
Reset criteria
The SpO
2
value goes back below the alarm limit.
Ventilator action The Nippy 4 will continue treatment according to the current
settings.
Setting range
90 % to 100 %
Off
Setting resolution 1 %
This alarm requires a connected SpO
2
sensor.
Alarms
93
Doc. 006553 M-6 User Manual