8.2.4 Monitored Values Specifications
This section describes the ranges and tolerances for monitored values on the Nippy 4.
All stated tolerances include measurement uncertainty. The accuracies have been tested with
all allowed configurations. Stated tolerances only disclose the maximum tolerance.
P
peak
Range/Performance: 4 to 99 cmH
2
O.
Tolerance: ±0.5 cmH
2
O or ±10%, whichever is greatest
EPAP
Range/Performance: 0 to 99 cmH
2
O.
Tolerance: ±0.5 cmH
2
O or ±10%, whichever is greatest
P
mean
Range/Performance: 0 to 99 cmH
2
O.
Tolerance: ±0.5 cmH
2
O or ±10%, whichever is greatest
CPAP
pressure
Range/Performance: 0 to 99 cmH
2
O.
Tolerance: ± (4% CPAP set pressure + 0.8 cmH
2
O)
Leakage
Range/Performance: 0 to 99.9 l/min (BTPS*).
Tolerance: ±10%
MV
Range/Performance: 0 to 99.9 l (BTPS*).
Tolerance: ±10% or (±10 ml × bpm), whichever is greatest
Vt
Range/Performance: 0 to 9999 ml (BTPS*).
Tolerance: ±10% or ±10 ml, whichever is greatest
FiO
2
Range/Performance: 0 to 100%.
Tolerance: ±2%
% in TgV
Range/Performance: 0 to 100%.
Tolerance: ±1%
Total Rate
Range/Performance: 0 to 99 bpm.
Tolerance: ±1 bpm