WARNING!
Risk of Electric Shock
Modifying or using the ventilator with accessories that are not specified or approved by
Breas may cause cardiac arrhythmia.
The Nippy 4 must only be used in original and unmodified shape and only with accesso-
ries specified or approved by Breas Medical.
Inadequate use of device or accessories may cause loss of treatment or decreased
performance.
CAUTION!
If you suspect that the device has been mistreated, perform a functional check before
taking it to use. A basic functional check can be performed as described in 4.5 Inspecting
the Nippy 4 before Use , page 39. A complete functional check can be performed by an
authorized service technician.
NOTE
Any serious incident that has occurred in relation to this device should be reported to
the competent authority and the manufacturer.
2.2 Electrical Safety
WARNING!
Risk of Electric Shock
High voltage contact may cause cardiac arrhythmia.
• Do not operate the Nippy 4 if it has a damaged power cord, power supply or
casing.
• To avoid electrical shock, only clean the Nippy 4 according to instructions in this
manual. Do not soak or immerse the Nippy 4 into any fluids.
• Use the approved power supply units only.
Use of unapproved power supply units may compromise the electrical isolation
and lead to risk of electric shock.
• Do not use more than one multiple portable socket-outlet or extension cord.
If a multiple portable socket-outlet is used, it must not be placed on the floor.
• The operator must not touch accessible contacts of connectors and the patient
simultaneously.
• Nurse Call must only be connected to a safety extra low voltage system with an iso-
lation from AC power (Mains) voltage which complies with the requirements of
IEC 60601-1.