Property Description
Alarm text
Disconnection
Priority High
Alarm condition A Disconnection alarm will be given when the measured flow
exceeds the expected leakage flow at the set Pressure for
15 seconds.
Possible cause
Too high leakage in the patient circuit.
The patient has removed the mask.
Circuit disconnection.
Reset criteria
The leakage is back within limits.
Ventilator action The Nippy 4 will continue treatment according to the current
settings
Setting range
On
Off
6.3.15 Rebreathing Alarm
Property Description
Alarm text
Rebreathing
Priority High
Alarm condition Leakage Circuit
A Rebreathing alarm will be given if the intentional leakage is
too low for more than 15 seconds.
Possible cause
Obstructed or occluded patient circuit.
Incorrect patient circuit.
Obstructed or removed CO
2
port from leakage circuit.
Reset criteria
The leakage is back within limits.
Ventilator action The Nippy 4 will continue treatment according to the current
settings.
Setting range
On
Off
Alarms
91
Doc. 006553 M-6 User Manual