6.4.19 High Humidifier Temp. Alarm
Property Description
Alarm text
High Humidifier Temp.
Alarm condition A Humidifier High Temperature alarm will be given if the
humidifier heater plate temperature exceeds 76°C (169°F) for
more than 2 seconds.
Priority
Medium
Ventilator action The ventilator will turn off the click-in humidifier and then con-
tinue treatment with the same settings.
A message with option to turn on the humidifier again will be
displayed.
Reset
The alarm is dismissed when the humidifier temperature drops
below 76°C (169°F), set humidifier temperature).
6.4.20 Humidifier Fault Alarm
Property Description
Alarm text
Humidifier Fault
Alarm condition
• All humidifier enabling conditions have been satisfied for 10
minutes, and
• No humidifier setting changes have been made for 10 minutes,
and
• Heater plate temperature < 50°C (122 °F)
• Humidifier set temperature > Ambient temperature ,and
• The heater plate temperature is more than 5°C (41 °F) below
the set temperature, or the heater plate temperature < -20°C
(68 °F) or greater than 400°C (752 °F)
Priority
Medium
Ventilator action The ventilator will turn off the humidifier and continue treat-
ment with the same settings.
The humidifier must be restarted manually when the cause of
the alarm is resolved.
Alarms
103
Doc. 006553 M-6 User Manual