6.4.23 Air Temp. Sensor Fail Alarm
Property Description
Alarm text
Air Temp Sensor Fail
Alarm condition The alarm is given in case of swivel boot temperature sensor
communication failure or sensor reporting temperatures out of
range (below -30°C (-22°F) or above 60°C (140°F).
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
6.4.24 Internal Error Alarm
Property Description
Alarm text
Internal Error
Priority High
Alarm Condition An internal Error alarm will be given when the ventilator has an
internal error, followed by an error code for the specific failure.
All Internal Error alarm error codes are defined and explained
in the ventilator Service Manual.
Ventilator action The ventilator will continue treatment with the same settings.
Reset action
Power off and restart the ventilator.
6.4.25 Database Integrity Fail Alarm
Property Description
Alarm text
Database Integrity Failed
Priority High
Alarm Condition This alarm is given when the database integrity check fails.
Ventilator action The ventilator will continue treatment with the same settings.
Reset action Rebuild the database and restart the ventilator.