6.4.14 LED Failure Alarm
Property Description
Alarm text
LED Failure
Alarm condition A LED Failure alarm will be given when one or more LED
indicators on the front panel are broken.
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
Reset
Power-on reset of ventilator (or repair).
6.4.15 Low Alarm Battery Alarm
Property Description
Alarm text
Low Alarm Battery
Alarm condition An alarm for Low Alarm Battery will be given if the alarm battery
is not charged enough to have power for a Power Fail alarm for
at least 2 minutes.
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings
and start charging the alarm batteries.
Reset
When alarm energy storage level is sufficient to give an alarm
for at least 2 minutes.
6.4.16 Alarm Battery Error Alarm
Property Description
Alarm text
Alarm Battery Error
Alarm condition Unable to communicate with super capacitor and read super
capacitor status.
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
Reset
When triggering condition is removed.
Alarms
101
Doc. 006553 M-6 User Manual