6.4.12 Temperature Comp. Lost (Ambient Temperature Compensation Lost
Alarm)
Property Description
Alarm text
Temperature Comp. Lost
Alarm condition An Ambient Temperature Compensation Lost alarm will be
given when the automatic ambient temperature compensation is
out of order.
There is no communication with the air temperature sensor or
the value is out of range (less than -30°C (-22°F) or more than
70°C (158°F).
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
The accuracy of the volume measurement may be impaired.
Reset
Ambient temperature inside valid range.
6.4.13 Humidity Comp. Lost (Humidity Compensation Lost Alarm)
Property Description
Alarm text
Humidity Comp. Lost
Alarm condition An Humidity Compensation Lost alarm will be given when the
automatic humidity compensation is out of order.
50% relative humidity is used for temporary compensation. If
the ventilator is used at other humidities, delivered and meas-
ured pressure and flow may deviate.
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
The accuracy of the volume measurement may be impaired.
Reset
Air humidity sensor values (RH and temperature) inside valid
range.
100
Alarms