8.3.5 Recommended separation distances between external power
conductors and the ventilator
Rated maximum current in
conductor (A)
Separation distance (m)
50-60 Hz d= I/2πH= I/188
1 0.005
10 0.05
30 0.16
For conductors rated at a maximum current not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation d=I/2πH, where I is the maxi-
mum current rating of the conductor in amperes (A) according to the transmitter manufac-
turer; H is the ventilator immunity compliance level to electromagnetic fields in the 50-
60 Hz frequency span (30 A/m).
8.4 Delivery Settings
Delivery settings: modes and functions
Ventilation Mode: Pressure Support
Device Mode: Clinical
Profile 1: Active
Profile 2: Off
Profile 3: Off