4 Reinstall the hatch.
5 Calibrate the FiO
2
sensor in the advanced settings of the main menu.
When installed, the ventilator automatically detects the sensor, also after powering off/
on and after power failure.
5.12.3.2 Calibrating the FiO
2
sensor
The FiO
2
sensor should be calibrated when first used and then at least once a month.
FiO
2
calibration can be performed from the “FiO
2
Calibration” page under the
Advanced settings section of the main menu.
5.12.4 Using the ventilator with the Remote Alarm
Information about safety, warnings, product description, installation, usage, cleaning,
maintenance and technical specifications can be found in the user instruction for Remote
Alarm.
The Remote Alarm enables care providers and clinical personnel to monitor the ventilator
alarms remotely. The Remote Alarm forwards alarms from the ventilator. When an alarm
sounds, the care provider or clinical personnel must attend to the patient quickly.
When installing a remote alarm system, check that it operates as intended before starting the
treatment.
5.12.5 Using the Ventilator with the SpO
2
module
Information about safety, warnings, product description, installation, usage, cleaning,
maintenance and technical specifications can be found in the user instruction for SpO
2
module.
The SpO
2
module enables connection to an SpO
2
sensor for measuring of functional oxy-
gen saturation of arterial haemoglobin (SpO
2
) and pulse rate. The SpO
2
module can be con-
nected to the Nippy 4 in order to monitor and store SpO
2
measurements.
The SpO
2
measurements will be stored in the data memory which can be downloaded to a
PC and viewed in the Breas PC software.
When installed, the Nippy 4 automatically detects the sensor, also after powering off/on
and after power failure.
How to Use the Nippy 4
69
Doc. 006553 M-6 User Manual