Usage
Time
Operating temperature
10 to 40°C (50 to 104°F)
Operating pressure
700 to 1250 mbar
Expected operating life <3 years (in ambient air) or 500,000 Vol.% h, which-
ever comes first.
Shelf life < 6 months (recommended)
CAUTION!
Note that the operating conditions for the FiO
2
sensor is different from the ventilator
system conditions. If the sensor is used outside its operating conditions the FiO
2
meas-
urements might deviate.
5.12.3.1 Installing the FiO
2
Sensor
1 Place the ventilator so the bottom is accessible.
2 Remove the hatch for FiO
2
sensor. Use
a torx TX10 screwdriver.
3 Insert the FiO
2
sensor with the electrical
contact side in.
68