6.3.12 Low Breath Rate Alarm
Property Description
Alarm text
Low Breath Rate
Priority High
Alarm condition A Low Breath Rate alarm will be given when the delivered total
breath rate is below the alarm limit for 15 seconds.
Possible cause
Mismatch between the Breath Rate setting and the alarm
setting.
The patient cannot trigger breaths because the inspiratory
trigger setting is too high.
Decrease in the patient’s spontaneous breathing.
Circuit disconnection.
Reset criteria
The breath rate goes above the alarm limit.
Ventilator action The Nippy 4 will continue treatment according to the current
settings.
Setting range
4 bpm to 40 bpm
Off
Setting resolution 1 bpm.
Alarms
89
Doc. 006553 M-6 User Manual