Setting Unit Min Max Default Resolution
Tolerance
Pressure Limit cmH
2
O Current
EPAP Max
+2
40 High Pressure
Alarm-2
0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
Relax Time Minute 2 12
Off
5 1
± (20 ms + 5% of set-
ting) or ±0.1 s,
Ramp Minute 10
Off
60
Off
10
± (20 ms + 5% of set-
ting) or ±0.1 s,
Always Start Ramp
-
Off
On
Off
- -
Ramp Start Pressure cmH
2
O 2 Current
IPAP–2
5 0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
Humidifier Setting Step 1 5 3 1
-
Heated Circuit Temp
°C/°F 16/61 30/86 27/81
0.5
Audible Alarm Level
Step 1 5 3 1 -
(1)= 0.5 cmH
2
O < 10 cmH
2
O , 1.0 cmH
2
O > 10 cmH
2
O, (3)= Also limited by IPAP- 2 cmH
2
O, (4)= breath per minute, (5)= Off is only avail-
able in Control mode, 6)Values >10 might initially be set with a 0.5 step when turning on Auto-EPAP. When changing the value, whole num-
bers will be used.
120
Technical Specifications
User Manual
Doc. 006553 M-6