Setting Unit Min Max Default Resolution
Tolerance
Min Pressure cmH
2
O 4 Current Max
Pressure
15 0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
Target Volume ml Off
300
(A)
50
(P)
2000
(A)
500
(P)
Off 10 <500
50 ≥500
±12 ml or ±10%,
Sigh
-
Off
On
Off
- -
Sigh Rate
1/Breath
10 250 50 10
-
Sigh Inspiratory Time
s
Current
Inpspira-
tory Time or
Backup Inspira-
tory Time
5 1.5 0.1
± (20 ms + 5% of set-
ting) or ±0.1 s,
Auto-EPAP
-
Off
On
Off
- -
EPAP Min
cmH
2
O 2 20 or Current
EPAP Max
5 0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
EPAP Max
cmH
2
O 2 or Current
EPAP Min
20 or Pressure
Limit-2
5 0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
EPAP Step
cmH
2
O 0.5 2 1 0.5
±0.5 cmH
2
O or ±5%
PS cmH
2
O 2 40–Current
EPAP Max
10 0.5 < 10
1.0 > 10
(6)
±0.5 cmH
2
O or ±5%
Min PS cmH
2
O 2 40–Current
EPAP Max
Variable
(1)
0.5 < 10
1.0 > 10
(6)
±0.5 cmH
2
O or ±5%
Max PS cmH
2
O 2 40–Current
EPAP Max
10 0.5 < 10
1.0 > 10
(6)
±0.5 cmH
2
O or ±5%
(1)= 0.5 cmH
2
O < 10 cmH
2
O , 1.0 cmH
2
O > 10 cmH
2
O, (3)= Also limited by IPAP- 2 cmH
2
O, (4)= breath per minute, (5)= Off is only avail-
able in Control mode, 6)Values >10 might initially be set with a 0.5 step when turning on Auto-EPAP. When changing the value, whole num-
bers will be used.
Technical Specifications
119
Doc. 006553 M-6 User Manual