Setting Unit Min Max Default Resolution
Tolerance
IPAP
cmH
2
O 4 40 15 0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
CPAP
cmH
2
O 4 20 10 0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
EPAP
cmH
2
O 2
Off
20
(3)
5 0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
Breath Rate bpm
(4)
4 50 12 1
±2%
Backup Rate bpm
(4)
4
0 (MPV)
50
40 (MPV)
12
0 (MPV)
1
±2%
Backup Insp. Time
s
0.3 5 1.5 0.1
± (20 ms + 5% of set-
ting) or ±0.1 s,
Inspiratory. Time
s 0.3 5 1.5 0.1
± (20 ms + 5% of set-
ting) or ±0.1 s,
Min Insp. Time
s
0.3
Off
3
Off
0.1
± (20 ms + 5% of set-
ting) or ±0.1 s,
Max Insp. Time
s
0.3 5
Off
Off 0.1
± (20 ms + 5% of set-
ting) or ±0.1 s,
Inspiratory Trigger Step 1 9
Off
(5)
3 1 -
Expiratory Trigger Step 1 9
(5)
3 1
-
Rise Time Step 1 9 3 1
-
Max Pressure cmH
2
O Current Min
Pressure
40 15 0.5 < 10
1.0 > 10
±0.5 cmH
2
O or ±5%
(1)= 0.5 cmH
2
O < 10 cmH
2
O , 1.0 cmH
2
O > 10 cmH
2
O, (3)= Also limited by IPAP- 2 cmH
2
O, (4)= breath per minute, (5)= Off is only avail-
able in Control mode, 6)Values >10 might initially be set with a 0.5 step when turning on Auto-EPAP. When changing the value, whole num-
bers will be used.
118
Technical Specifications
User Manual
Doc. 006553 M-6