EPAP 5 cmH
2
O
Rise Time 9
Insp. Trigger 9
Exp. Trigger 3
Min Insp. Time Off
Max Insp. Time Off
Backup Rate 12 bpm
Backup Insp. Time
2.0 s
Target Volume Off
5 All alarm settings shall be set to Off if possible.
6 Start the treatment.
6.5.2.1 High and Low Flow Alarm Tests
1 Set the high flow alarm to 20 l/min.
⇒ The high flow alarm shall be given.
2 Set the high flow alarm to Off
3 Set the low flow alarm to 150L/min.
⇒ The low flow alarm shall be given.
6.5.2.2 High and Low Pressure Alarm Tests
1 Set the high pressure alarm to 10 cmH
2
O.
⇒ The high pressure alarm shall be given.
2 Set the high pressure alarm to 55 cmH
2
O.
3 Set the low pressure alarm to 20 cmH
2
O.
⇒ The low pressure alarm shall be given.
4 Set the low pressure alarm to 1.0 cmH
2
O.
6.5.2.3 Expiratory Tidal Volume Alarm (Vt
e
) Tests
1 Set up the ventilator as described in Alarm Test Preparation, page 107.
2 Set the high Vt
e
alarm to 150 ml.
⇒ The high Vt
e
alarm shall be given.
3 Set the high Vt
e
alarm to Off.
4 Set the low Vt
e
alarm to 400 ml.
The low Vt
e
alarm shall be given.
6.5.2.4 SpO
2
Related Alarm Tests
These tests applies if the SpO
2
accessory is used,
108
Alarms