Configuration
7.4.5.3 Calibrating the Thermocouple
A 2 point correction (e.g. Ice water & Liquid N2) of the thermocouple reading can be
done under:
Main -> VTU -> Thermocouple
7.4.6 Peripheral Control
Since peripheral control is a slave of the VTU, the specific web server functions are
found on the VTU main board web server under:
Main-> VTU -> Peripheral Control