Z31980_00_03
Installation
Newer models just have two external filters (E1451420, E1451430)
• E1451420 is connected to 1H Connector the rear of the console.  
• E1451430 is connected to C13 Connector the rear of the console.  
6.3 Adjusting Air Pressure
Adjust the air pressure so that the manometer in the console reads between 3.5 and 4 
bar. That way the default spin and lift parameters will get you started.
6.4 Switch on for the first time
The console has three modes:
• OFF (Blue light OFF), switch at rear panel set to OFF.
• STAND BY (Blue light BLINKING), switch at rear panel set to OFF, ON/OFF Button 
at top of console not yet pressed.
• ON (Blue light ON). 
i
Note: It is IMPORTANT to switch the console on BEFORE switching on the PC, since 
the console contains the Ethernet hub, and the DHCP Server will not start if it doesn't 
find a network. 
6.5 Checking that the Console has booted
This procedure takes 30-60 seconds, provided the DCHP Server is running. There is no 
visual indication on the console, but the easiest way to check is to see if the three main 
Ethernet devices (Main Board, Lock and VTU) can be reached via ping or a web 
browser. Their addresses will usually be in the range:
149.236.99.25x
If you are not sure the DHCP Server is running, you can check by opening
Control Panel -> Administrative Tools -> Services -> Bruker DCHP Server