16
Z31980_00_03
Safety
Impacting Magnet
Genuine Samples
3.4.2 Dangers from Electric Power
Stored Charges
NOTICE
Material damage hazard due to impacting the magnet!
Impacting the magnet may result in a quench.
Mount the device carefully on the magnet.
Avoid banging the magnet during installation and operation, e.g. when replacing
the sample cassette.
NOTICE
Material damage due to the use of genuine samples during installation
and maintenance!
Using genuine samples during installation and maintenance may result in material
damage.
Use only dummy samples during installation and maintenance.
DANGER
Danger to life from stored charges!
Electric charges may be stored in electrical components even after the system has
been switched off and disconnected from the power supply. Contact with these com-
ponents may result in serious or fatal injury.
Before working on the specified components, ensure that they have been com-
pletely disconnected from the power supply. Allow 10 minutes to elapse in order to
ensure that the internal capacitors have been fully discharged.

Table of Contents

Other manuals for Bruker Fourier 300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker Fourier 300 and is the answer not in the manual?

Bruker Fourier 300 Specifications

General IconGeneral
BrandBruker
ModelFourier 300
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals