Installation
For the ATP Tests you will need the following samples
The three Test Samples which were shipped with the system.
3% CDCl3 in Chloroform (lineshape)
0.1% EB in CDCl3 (1H SINO)
ASTM (C13 SINO)
To shim the magnet and perform the ATPs test you will require to bring some additional
samples:
Urea in DMSO (Pulse Length adjustment)
100mg or 1g Cholac in CDCl3 (DEPT)
Cyclosporine in C6D6 (2D Examples)
There are only three hard specs for the Fourier required for a customer sign off:
The remaining ATP Tests will implicitly demonstrate the system remaining functions:
Water Supression
Sample Spin & Lift
Auto-Locking on various solvents
Gradient Functionality (2Ds & Gradshim)
Gradshim reproducibility
Lock and Temperature Stability (COSY & HSCQ)
* Baseline & Spectral Purity (DEPT)
* Certain Artefacts are normal and have been considered acceptable for the
intended use of the Fourier:
A moderate amount of T1 noise in 2D experiments
Some "digital noise" on strong signals that require RG < 10. Use zg30 instead zg of
if that happens, in order to be able to increase RG
Some vibration noise close to the feet of very narrow lines (e.g. lineshape samples).
This can be improved with vibration damping if required.
* Some very small spurious spikes in the first few scans of C13 spectra. These
should average out after NS > 64
Lineshape using 3% CDCl3 in Acetone 0.5/8/16 Hz (non-spinning)
1H Sensitivity (0.1% EB in CDCL3) 125:1 (after gradshim)
C13 Sensitivity (ASTM) 90:1 (after gradshim)
Table 6.1 Hard specs for the Fourier