48
Z31980_00_03
Installation
binding state active;
next binding state free;
hardware ethernet 00:00:ad:12:85:12;
uid "\001\000\000\255\022\205\022";
Once you are sure that all three devices have booted, you can start TopSpin. Normally
your PC will come with CF having already been carried out. If you get an error message,
that one or more devices cannot be found, you may have to repeat CF.
6.6 Performing cf on a Fourier
TopSpin 3.1 is the first software to be used with the Avance system and the Fourier sys-
tem. All changes to TopSpin regarding the new Fourier console are implemented in the
same software tree. That means that a Fourier spectrometer can be controlled from any
computer running TopSpin 3.1.
As with the Avance systems, the Fourier must be configured in TopSpin using the cf
command. The NMR administrator password is required to run cf.
The first screen allows to edit an already existing configuration or to create a new one.