7.1.1 Resetting the CPU
Any change is network configuration, firmware of FPA content are only valid after
rebooting the CPU. There are three ways to reset the CPU:
Power up
Pressing the reset button on the CPU
Pressing “Hardware reset” button in the web server under
Main -> support -> network configuration
This button also appears after a successful firmware update.
7.2 Firmware Update
The firmware of all the intelligent devices can be upgraded during run-time via the web-
browser. The mechanism for the Ethernet Devices differs to that of the CAN-Bus
Devices.
7.2.1 Ethernet Devices (Main Board, Lock Board, VTU)
Downloading a new firmware is initiated by:
Main -> support -> CPU software update
Before changing Firmware the
current firmware version must be
identified. In Ethernet Devices
this is always shown at the bot-
tom of each web server page.
1. Select the firmware to be
uploaded using the “browse
button (in some versions
called “Durchsuchen”.
2. Press Upload to upload firm-
ware to CPU.