EasyManuals Logo

Bruker Fourier 300 User Manual

Bruker Fourier 300
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
114
Z31980_00_03
Configuration
7.5 Additional Test Applications
The Fourier has some additional test applications the route certain RF signals back to
the receiver in order to check if all the RF paths are working. These tests are based on
the minispec software which is routinely installed with Fourier PCs.
The minispec Fourier Test Applications are found under:
C:\program files (x86)\Bruker the minispec\service_only\fourier test applications
FID.app
acquires an fid independently from topspin. Pulse length & frequency etc can be
adjusted under "settings". TD, SWH are fixed for this test.