Configuration
7.4.1.3 On-Board Voltages
The main board receives only 28Vdc from the central 28V PSU and generates all the
other voltages it requires on board. You can test if all the voltages are correct under:
Main-> minispec-> AD Controller
7.4.1.4 4.5.1.5 Pulse programmer memory
Unlike the IPSO, the Fourier keeps its entire Pulse Program (incl. decoupling
sequences, loops, 2D evolutions) in memory at once. Since the memory is limited, some
pulse programs may exceed the memory capacity. In order to check if this is the case,
you can check the memory usage using:
Main-> minispec-> memory