89
Z31980_00_03
Configuration
the embedded CPU. The PID settings can be adjusted using:
Main-> minispec-> electronics temperature control
A temperature setpoint of 1.0 is equivalent to about 40°C. It should be adjusted so that
the heater power is between 30 and 50 %. when stable conditions are reached. Adjust
the setpoint if necessary (lower value -> higher temperature. The setpoint is raised by
1°C when the setpoint value is lowered by 0.05, and vice versa ).
PID Values should not need to be changed. The values are stored in the EEPROM and
are valid after power up.