67
Z31980_00_03
Installation
the lock shift. The new shift value is set in the Fourier and a normal lock procedure is
started.
Figure 6.2 Result of Flock with one Solvent Peak
Result of flock if number of detected peaks is not compatible with number of
expected solvent peaks. This can happen if a mixture of solvents is used or if the
wrong solvent has been chosen. In any case, it leads to a field lock. The ppm scale
may be shifted if the wrong solvent has been chosen, though.
Any time, the lock field has changed by a large amount, the following message is shown.