50
Z31980_00_03
Installation
6.6.2 ETH Addresses and tty Settings
During the cf run, some information is shown about the detected hardware. In addition to
the Fourier main board, only the lock system, the VTU and the Lockswitch are part of the
basic instrument. The Ethernet addresses of the Lock and the VTU are set by the DHCP
server and don’t need to be set manually. A missing entry shows that the respective
board couldn’t be detected. This may indicate a problem of the board or the DHCP
server.
The Lockswitch has to be set to amplifier 1, which is the X amplifier.
The preamplifiers and power amplifiers are build-in and are directly controlled by the
Fourier main board.
Currently, none of the Avance accessories is used with the Fourier system. This may be
done in the future, though.
6.6.3 Cortab and Power Check
Cortab files for the console are already measured as part of the production. So, please
enable power check for probe protection.
6.6.4 Editing the Nuclei List
It might be a good idea to edit the nuclei list. The Fourier is a dedicated 1H, 13C and 2H
instrument. So, only these nuclei are needed. This is not mandatory but a lot of selection
windows are much easier to scroll if the nuclei list has been cut short.