EasyManua.ls Logo

Bruker Fourier 300 - Setting the Spin Safety Threshold

Bruker Fourier 300
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
110
Z31980_00_03
Configuration
7.4.6.3 Setting the Spin Safety Threshold
The spin has a safety function that shuts the spin off as soon as the Lift is turned on
either manually or during the initialisation of the sample changer. In order to do this, the
analog sensor that monitors the spinner has to be calibrated. It has three voltages for the
three possible states:
Since the values vary between shim stacks and also depends somewhat on the spinner
type, it is necessary to calibrate this value. Usually this only has to be done once at
No spinner (typically 0.1 to 0.6V)
Black (typically 0.6 to 2.5V)
White (typically 2.5 to 4.5 V)

Table of Contents

Other manuals for Bruker Fourier 300

Related product manuals