7-20 User Manual
Message Reason for Appearance
40-03 Recalled Needle
This needle lot number is identied as part of a recall
and is unavailable for use
Return the needle to Galil Medical
Connect a needle from a di󰀨erent lot number to
continue the procedure
The needle lot number was identied as
a lot number in a regulatory recall. The
channel will be disabled until a needle
from a di󰀨erent lot number is connected.
40-04 Expired Needle
The needle ‘Use by’ date is in the past
Replace with a new needle
The needle expiration date was identied
as an expired date. The channel will be
disabled until an appropriate needle is
connected.
40-05 Used Needle
This needle has been previously used
Replace with a new needle
The software prevents needle reuse
and identied a needle previously used.
The channel will be disabled until a new
needle is connected.
40-06 Corrupt Memory
Needle Memory chip is corrupt on Channel X
Manually select the needle type
User connected a needle containing
a bad memory chip. User may select
needle type from the drop-down menu.
40-07 Passive Thaw Required
The currently selected needle on channel [x] does
not have i-Thaw capability
Since helium is not connected, passive thaw is
required on this channel
Combinations of needles that are i-Thaw
capable and that are not i-Thaw capable
were connected to the needle connection
panel; no helium gas was connected, so
passive thaw was required.