7-4 User Manual
Symptom Potential Causes / Solutions
The Login Screen appeared
after the system was left idle for
more than 2 hours while on the
Procedure Screen
Enter the appropriate password to return to the Procedure
Screen.
Touch Screen goes blank during a
procedure
The video cable may be disconnected.
WARNING. Do not touch the screen if the touch-screen
monitor goes blank for greater than ve (5) seconds
during a procedure. Immediately turn o󰀨 power to the
system and end the procedure to avoid inadvertent
activation of needles.
Contact Galil Medical Customer Service.
7.2.1 Replacing Fuses
Instruction
Photo
1. Turn the Power Switch OFF on the rear of the Visual-
ICE System. Loosen the thumb screw on the cord
retention clip.
2. Remove the power cord from the retention clip.
Remove the two screws that secure the retention clip
and remove the retention clip from the power inlet.