Visual-ICE™ Cryoablation System
7-5
3. Insert a small screwdriver into the slot on the bottom
of the fuse holder to begin to snap the fuse holder
out of the power inlet.
4. While holding your hand under the fuse holder,
carefully slide the fuse holder out of the power inlet.
There are four fuses in the fuse holder.
5. Catch the fuse holder and fuses as the fuse holder
is removed from the power inlet. The two fuses that
remain in the fuse holder are the fuses that are part
of the system circuit.
6. Replace the fuses in the fuse holder with the
two loose fuses. NOTE: Use only Galil Medical
specied fuses in the Visual-ICE System.