Visual-ICE™ Cryoablation System
8-1
8 SYSTEM SPECIFICATIONS
If additional technical information is required, place a request with a Galil Medical qualied service
engineer.
8.1 Operating Conditions
Temperature: 10°C to 40°C
Relative Humidity: 30% to 75%
8.2 Storage Conditions
Temperature: -15°C to + 50°C
Relative Humidity: 10% to 90%
8.3 Transportation Conditions
When shipping the Visual-ICE™ Cryoablation System, use the original shipping containers to prevent
damage from occurring during transport. If the original shipping containers are not available, the
customer takes responsibility to ensure proper transport conditions are satised or contacts Galil
Medical Customer Service to obtain the appropriate shipping container.
8.4 Mechanical Specications
Weight: 100 kg (220 lbs)
Height: 107 cm (42 in), monitor down
157 cm (62 in), monitor up
Foot Print: 56 x 66 cm (22 x 26 in)
Storage Compartment Weight Capacity: 22 kg (50 lbs)
Monitor Storage Basin Weight Capacity: 9 kg (20 lbs)
Closed Monitor Weight Capacity: 9 kg (20 lbs)
8.5 External Gas Supply
Argon Cylinder:
o Purity Level: 99.998% or higher
o Solid particle size: < 5 µm
Helium Cylinder:
o Purity Level: 99.995% or higher
o Solid particle size: < 5 µm