EasyManua.ls Logo

BTG Visual-ICE - 4 System Operation; Preparation for Use; Table 4-1. Cryoablation Procedure Flow

BTG Visual-ICE
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Visual-ICE™ Cryoablation System
4-1
4 SYSTEM OPERATION
Table 4-1 depicts the order and steps of the Visual-ICE™ Cryoablation System set-up and treatment
procedure. Each step is described in detail in this chapter.
Table 4-1. Cryoablation Procedure Flow
1.
Set-up Console
Conrm availability of gas, needles, and accessories
Position Visual-ICE System and lock brake
Ensure Manual Vent Valve is closed and Argon
Shuto󰀨 Valve is ON
Power Visual-ICE System ON
Login
2.
Connect Gas
Cylinders
Connect helium gas, if applicable
Connect argon gas
Ensure safety cables are connected
OPEN valve on helium, then argon
3.
Needle / MTS
Testing
Enter patient treatment information in Registration
Button
Prepare sterile needles for testing
Connect needles to Visual-ICE System and lock
channels
Perform Needle Integrity and Functionality Test
Test Multi-Point Thermal Sensors, if applicable
4.
Conduct
Cryoablation
Procedure
Insert needles into target tissue
Deliver the cryoablation treatment and remove
needles
View and save reports, if desired
Depressurize the system and shutdown
4.1 Preparation for Use
Prior to using the Visual-ICE System, inspect the chassis, power cord, brake, safety cables, gas
supply lines, gas connections, and the monitor touch screen to ensure they are not damaged. If any
of the components are damaged, contact Galil Medical Customer Service.
CAUTION. If condensation is present on the system, allow the system to dry completely for 12
hours prior to powering up the system. Powering up the system with condensation could result in
permanent damage to the electrical boards, causing the system to be inoperable.
Before the patient is anesthetized, set up the Visual-ICE System, connect the gas cylinders, and
conduct functionality tests on each cryoablation needle and thermal sensor (reference Section 4.1.4).

Table of Contents