4-10 User Manual
Screen 4-9. Optimizing Lines message
4.1.4 Pre-Procedure Testing
WARNING. Prior to beginning a treatment and before the patient is anesthetized, set up the
Visual-ICE System and perform functionality tests on each cryoablation needle and thermal
sensor.
1. On the touch screen monitor, press Start Procedure. The Procedure Screen appears (Screen
4-10).
Screen 4-10. Procedure Screen
2. Using aseptic technique, carefully remove the cryoablation needle from the package and place in
a sterile work area.
3. Remove the connector cap, and then connect the needle to the Visual-ICE System needle
connection panel (Figure 2-4).
CAUTION. Do not kink, pinch, cut or pull excessively on the needle tubing. Damage to needle handle
or tubing may cause the needle to become unusable.