EasyManua.ls Logo

BTG Visual-ICE - Figure 4-3. EZ-Connect2 Dual Cylinder Adapter

BTG Visual-ICE
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-8 User Manual
• Connect the auxiliary gas supply line to the four-way adapter assembly using the quick-
connect connector located on the end of the auxiliary gas supply line.
• Connect the opposite end of the auxiliary gas supply line to the second argon cylinder by
securing the end of the auxiliary line to the cylinder connection.
• Open the primary cylinder valve rst and use this cylinder until depleted. Do not open the
cylinder valve on the second cylinder until the initial cylinder has been depleted.
• Reference Section 4.5 for instructions on changing the gas cylinder during a procedure, if the
second cylinder is also depleted during the procedure.
2
3
1
Figure 4-3. EZ-Connect2 Dual Cylinder Adapter
1 Auxiliary Gas Supply Line
2
Four-way Adapter Assembly with
Pressure Gauge
3 Gas Supply Line
CAUTION. Before beginning the cryoablation procedure, ensure availability of su󰀩cient gas to
conduct the planned cryoablation procedure (reference Section 8.5 for gas purity requirements).
10. Ensure the Gas Indicator (Screen 3-4) displays the minimum working pressure before initiating a
procedure (Table 4-2). The Gas Indicator should indicate the pressure to be in the green range.
If the system detects that the pressure reading for either gas cylinder is less than 50 psi, a
message displays on the Navigation Tool Bar (Screen 4-8). Connect the gas cylinders to the
Visual-ICE System.

Table of Contents