EasyManuals Logo

Cadex Electronic C7000-C Series User Manual

Cadex Electronic C7000-C Series
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #142 background imageLoading...
Page #142 background image
Cadex Electronics Inc. C7000-C Series Battery Analyzers .
126
Detailed
Message
Code Reasons Solutions
INSERT THE
BATTERY
8
Insert battery in the Battery
Adapter.
INTERMITTENT
BATTERY
129
Battery connection opened
five times in one minute.
Check and clean the battery
contacts. Verify that correct
contacts are used. Verify that the
battery is properly inserted.
Verify that the C-Code is correct
for the battery.
Reduce the charge rate in
extended C -Code. Increase end-
of-discharge voltage.
For intrinsically safe batteries,
use I/S settings (see
(Intrinsically Safe Batteries,
page 53).
I
NTERNAL
E
RROR
(
NONE
)
Display Buffer is full Machine will reset in a few
minutes a resume service. If
message persists, press the Esc
key for a few minutes, the
machine will reboot and resume
service.
This may occur if QuickTest (a
processing intensive program) is
used. Start QuickTest in each
station after 10 seconds.
INVALID PASSWORD
ENTERED
204
(Code 204) An incorrect
password has been entered.
Enter the correct password.
If correct password is unknown,
contact Cadex with unit serial
number.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cadex Electronic C7000-C Series and is the answer not in the manual?

Cadex Electronic C7000-C Series Specifications

General IconGeneral
BrandCadex Electronic
ModelC7000-C Series
CategoryMeasuring Instruments
LanguageEnglish