Cadex Electronics Inc. C7000-C Series Battery Analyzers .
Detailed
Message
Code Reasons Solutions
THERMISTOR
FAILURE
150
The battery temperature
reading is invalid.
Clean the contacts on the
battery.
Thermistors on non-OEM
batteries may not be the same
as that used on OEM batteries.
Contact Cadex to upgrade the
adapter. If the battery is NiCd or
NiMH, turn off the temperature
sensing.
If the battery is known to be
good or the adapter has two
contacts (in these adapters, the
negative contact is a thermistor),
the battery thermistor is
defective, contact Cadex.
TRICKLE CHARGE
3
The program is running a
trickle charge.
The reconditioning process is
complete, and the battery is
being recharged.
U
NABLE TO
L
EARN
M
ATRIX
179
A matrix cannot be created
because: battery capacity is
less than 50%, the OhmTest
reading does not produce a
result or the battery cannot
handle 1.00C charge or
discharge rate required by
Learn
Use another battery to run Q-
Learn or Learn.
USER PROGRAMMED
TIMEOUT
33
The time programmed in a
custom program has
completed.
The program goes to the next
step.