EasyManua.ls Logo

Circutor CVM-E3-MINI - MEASURING PARAMETERS

Circutor CVM-E3-MINI
124 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IEEE measurement convention
Q
P
Q1Q2
Q3
Q4
P < 0 Q > 0 PF > 0
Capacitive
Capacitive
Inductive
Inductive
P < 0 Q < 0 PF < 0 P > 0 Q < 0 PF > 0
P > 0 Q > 0 PF < 0
Operation in the 4 quadrants (Q1, Q2, Q3, Q4)
0 -
0 +
cos φ < 0
cos φ > 0
Q1
Q4
- 1
+ 1
cos φ values in the receiver operating mode (Q1,Q4)
Figure 25: IEEE measurement convention.
4.1.- MEASURING PARAMETERS
The device displays the electrical parameters shown in Table 11.
Table 11: Measuring parameters of the CVM-E3-MINI
Parameter Units
Phases
L1-L2-L3
Total
III
Maxim.
value
Minim.
value
Phase-neutral voltage Vph-N
Phase-phase voltage Vph-ph
Current A
Neutral Current
(1)
A
Frequency Hz
Active power M/kW
Apparent power M/kVA
Total Reactive Power M/kvar
Total Reactive Power - Consumption M/kvar
Total Reactive Power - Generation M/kvar
Total Inductive Reactive Power M/kvarL
Inductive Reactive Power - Consumption M/kvarL
Inductive Reactive Power - Generation M/kvarL
Total Capacitive Reactive Power M/kvarC
Capacitive Reactive Power - Consumption M/kvarC
Capacitive Reactive Power - Generation M/kvarC
Power factor PF
Cos φ
φ
THD % Voltage % THD V
THD % Current % THD A
34