Press the   key for 3 seconds to enter the edit mode. The prog icon will flash on the left of the 
screen.
Press the   key to modify the value of the flashing digit. 
When the desired value is shown on the screen, use the   and   keys to move the edit cursor.
To validate the value, press the   key for 3 seconds; the prog icon will stop flashing.
If the value entered by the user is out of the range of programming values, the programmed value will 
be deleted and the system will restore it to the last saved value.
 Configuration values
Table 22: Configuration values: Secondary voltage.
Secondary voltage
Minimum value 1 V
Maximum value
999 V
Voltage ratio ≤ 1000.
Voltage ratio x Current ratio ≤ 300000.
Press key   to access the next programming step.
 
 6.3.- PRIMARY CURRENT
Note: Display visible on models: CVM-E3-MINI-ITF, CVM-E3-MINI-ITF-WiEth, CVM-E3-MINI-MC and 
CVM-E3-MINI-MC-WiEth. 
The current transformer primary is programmed on this screen.
Press the   key for 3 seconds to enter the edit mode. The prog icon will flash on the left of the 
screen.
Press the   key to modify the value of the flashing digit.
When the desired value is shown on the screen, use the   and   keys to move the edit cursor.
 
To validate the value, press the   key for 3 seconds; the prog icon will stop flashing.
If the value entered by the user is out of the range of programming values, the programmed value will 
be deleted and the system will restore it to the last saved value.