Page #124 background imageLoading...
Page #124 background image
Performing a Test
124
With operator login (list available)
2 If an operator list exists, wait until this list
is displayed.
3 Touch or to scroll through the list.
Select the operator of choice by touching
the corresponding button.
As an alternative (meters with a scanner),
the operator login can be performed via the
built-in barcode scanner.
Touch Scan
Hold the card with the barcode approx.
15 cm (6 in) from the scanner.
It is not possible to connect an external bar-
code scanner to the meter.
The scanner equipped cobas h 232 meter
described in this manual provides a camera
based 2D barcode scanner, instead of the
laser scanner used in previous meter versions.
Established scanning procedures may need
some corresponding adjustments:
the nominal distance between scanner
and barcode is slightly different
avoid reflections on the barcode (e.g. by
reflective foils etc.)
ensure an adequate size and quality of
barcodes
See Appendix A for further details on barcode
handling.
OP003
Hall, Molly-Jessica
OP001
Hughes, Adam
OP002
Patel, Alfi e-Adam
Select Operator ID
Scan
04/19/2016

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen