Performing a Test
131
If you use a new test strip lot, you must insert
the corresponding code chip once.
See “Code chip” on page 118.
Depending on the meter setting, you may also
be required to run a quality control test at this
point.
The meter identifies the required code chip
based on the test strip barcode and displays
the code number.
For each test parameter and each new test strip lot a different code key number is displayed.
The thermometer icon shows that the test
strip is warming up. A beep tone indicates
when the warming-up process is complete
(provided the beeper is enabled).
Pat. Test - Ins. Chip
Insert Code
Chip
proBNPPar:
04/19/2016
2721
Pat. Test - Warming Up
Please Wait
Warming Up
PID111SCHULZM457
SCHULZ, MANFRED
JONES, TOM
Pat:
Op:
proBNPPar: PB1234Code:
04/19/2016

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)