Maintenance 7-7
5. The program analyses alignment and stitching status of the cameras. If the cameras
are unaligned, the program and scanner will automatically make the necessary
corrections.
6. After Camera Alignment, the Scanner Maintenance program automatically moves on to
Calibration.
7. Calibration completes the maintenance process. Remove the Scanner Calibration
Sheet from the scanner.
8. Return the Scanner Calibration Sheet to its protective cover and then place it in the
storage folder. Store the folder in a dry place and out of direct light.