6.3.9 End a run
Fo
llow the steps below to end a run.
Make sure that the syringe pushers have returned to their home positions and the
status on the user interface displays Completed.
Step Action
1 Open the lid of the instrument.
2 Remove the sample and waste collection tubes.
3 Rotate the rotating block upwards.
4 Remove the syringes.
5 Rotate the rotating block into the default position and remove the cartridge.
6 If applicable, remove the cartridge adapter.
7 If applicable, wait for the heating block to cool and remove the heating
in
serts.
8 Dispose of the used cartridge and syringes according to local regulations.
9 Clean any spills or drips immediately. See Section 7.4 Routine cleaning,
on page 109 for cleaning instructions.
Note:
When running a procedure on the Ignite+ instrument, some spillage is
expected during the procedure due to the high flow rates. The spillage is
minimized as much as possible by the system.
6 Operation
6.
3 Run a procedure
6.3.9 End a run
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 104