4.1.4 Power requirements
In
troduction
This section describes the electrical power requirements for the Ignite or Ignite+
instrument.
WARNING
Protective ground. The product must be connected to a
grounded power outlet.
WARNING
Po
wer cord. Only use power cords with plugs delivered or
approved by Cytiva.
Electrical power requirements
Th
e following table specifies the power requirements for the instrument.
Parameter Requirement
Supply voltage 100 to 240 V~
Supply current 4 A
Output voltage 24 Vdc
Output current 6.3 A
Output power 151 W
Frequency 50/60 Hz
Power outlet type Grounded mains outlets
Transient overvoltages Overvoltage category II
Location of sockets Maximum 2 m from the instrument (due to length of
ma
ins cable). Extension cables can be used if required.
Power quality
T
h
e mains power supply must be stable and conform to specifications at all times to
ensure reliable operation of the instrument. Avoid transient or slow changes in average
voltage outside the limits specified above.