Power failure
Th
e following table describes the consequences of a power failure.
Power
fa
ilure to...
will result in...
Ignite or
Ig
nite+
instrument
• The run is interrupted immediately and the instrument stops
f
ormulating.
Note:
The fluid already dispensed can be collected and saved.
• If heating was used, the power to the heaters is cut, but the
surfaces remain hot for some time.
CAUTION
Ho
t surface. Avoid direct contact with
the heating block and the top surface of
the rotating block during or after heating.
Contact with hot surfaces during or
directly after a procedure in which the
heaters are on can cause burns.
• The run has the status INCOMPLETE in the History menu.
Restart after emergency shutdown or
po
wer failure
Before restarting the instrument, make sure the power switch on the back of the
instrument is set to the O position. Follow the steps below to restart the instrument:
Step Action
1 If necessary, plug in the power cord to the wall socket.
2 Press the power switch to the I position to restart the instrument.
Result:
Upon restart, all of the system components return to their home positions.
3 Remove any syringes or collection tubes.
The instrument is ready to start a new formulation.
2 Safety instructions
2.
3 Emergency procedures
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 15