4.1.2 Room requirements
In
troduction
This section describes the requirements for the room where the system is placed.
Dimensions and weight
The following illustration shows the dimensions of the Ignite instrument.
Parameter Value
W 38 cm (15 in)
D 34 cm (13.4 in)
H 38 cm (15 in)
Weight 16 kg (35.3 lb)
Space requirements
WARNING
Ac
cess to power cord. Do not block access to the power cord. The
power cord must always be easy to disconnect.
CAUTION
Ma
ke sure that the system is placed on a stable, level bench with
adequate space for ventilation.
4 Installation
4.
1 Site preparation
4.1.2 Room requirements
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 45