4.1.1 Delivery, storage and unpacking
In
troduction
This section describes the requirements for receiving the delivery box and storing the
instrument before installation.
This section outlines important information that must be considered when handling
the transport crate, unpacking the system and moving the system after unpacking.
Two people are recommended to safely unpack and move the instrument, and no
special equipment is needed.
When you receive the delivery
Record on the receiving documents if there is any apparent damage on the delivery
box. Inform your Cytiva representative of such damage.
Move the delivery box to a protected location indoors.
Delivery box
Ignite and Ignite+ instruments are shipped in a delivery box with the following dimen-
sions and weight:
Contents Dimensions (W × D ×
H)
W
eight
Ignite or Ignite+ instrument
wi
th accessories
51 × 51 × 51 cm (20 ×
20 × 20 in)
22 kg (48.5 lb)
Storage requirements
Th
e delivery boxes should be stored at a protected place indoors. The storage place for
unopened boxes must meet the following requirements:
Parameter Allowed range
Ambient temperature, storage 15°C to 35°C
Relative humidity 25% to 65%, non-condensing
Transportation
T
h
e following equipment is recommended for handling the delivery boxes:
Equipment Specifications
Cart for transporting the
in
strument to the lab
Dimensioned to accommodate the size and
weight of the instrument
4 Installation
4.
1 Site preparation
4.1.1 Delivery, storage and unpacking