4.1.3 Site environmental requirements
In
troduction
This section describes the environmental requirements and conditions for installation
of Ignite or Ignite+.
CAUTION
The product is designed for indoor use only.
Environmental requirements
Th
e installation site must comply with the following specifications.
Parameter Requirement
Allowed location Indoor use only
Ambient temperature 15°C to 35°C
Relative humidity 25% to 65%, non-condensing
Pressure 0.070 to 0.106 MPa
Altitude Maximum 2000 m
Pollution degree of the intended envi-
ro
nment
Pollution degree 2
Environmental conditions
Th
e following general requirements must be fulfilled:
• The room must have exhaust ventilation.
• Dust in the atmosphere should be kept to a minimum.
Heat output
The heat output data is listed in the table below. The heat output data is valid when the
syringe heaters are installed and on.
Component Heat output
Syringe heaters on the Ignite or Ignite+
instrument
Maximum output per heater: 35 W
Total Maximum: 70 W
4 Installation
4.
1 Site preparation
4.1.3 Site environmental requirements
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 47