5.1.4 Parameter validation
Af
ter all run parameters have been entered, the parameter combination is validated. If
run parameters are missing or outside the operational limits of the instrument, an
error message is displayed in red at the bottom of the screen.
Examples of incompatible parameters include:
The combination of Total flow rate, Total volume, and Flow rate ratio causes the
a flow rate for a single channel to exceed the operational limit.
The combination of Total volume and Flow rate ratio causes the dispense volume
for a single syringe to be larger than the syringe size.
The Total flow rate is higher than the maximum operational flow rate.
For some validation errors, the problematic parameter is highlighted in red. For valida-
tion errors that are the result of calculated values for a single channel, no parameters is
highlighted, as there is more than one parameter value that can be adjusted to fix the
error.
When resolving validation errors, use the volume and flow rate calculations displayed
for each channel under the syringe drop-down lists.
If errors are present, adjust the run parameters to within the flow rate and volume limi-
tations. See Section 5.1.2 Mixing parameters, on page 59 for the range of acceptable
flow rates and volumes.
5 Preparation
5
.
1 Parameters
5.1.4 Parameter validation
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 67