NOTICE
Do
not use polycarbonate syringes for the solvent phase. Use a
syringe made from polypropylene instead.
Note:
Some syringes, particularly those that have an elastomeric plunger seal, are
ma
de of more than one fluid contacting material.
Collection tube compatibility
Ce
rtain collection tubes have been validated for use with the instruments, and the
system performance has been optimized for these products. The table below gives the
compatible tubes for sample collection and waste collection.
Tube type Compatible tubes
Waste collection
15 mL centrifuge tubes
Ignite sample collection
15 mL centrifuge tubes
5
0
mL centrifuge tubes
Ignite+ sample collection
15 mL centrifuge tubes
5
0
mL centrifuge tubes
100 mL Nalgene round-bottom centri-
fuge tubes
Note:
15 mL and 50 mL Falcon conical tubes have been validated for geometric
ch
aracteristics.
If you choose to use an alternate collection tube, consult the manufacturer of the
sa
mple collection tubes for chemical compatibility and the required level of cleanliness
of the external environment. Consider the mixture of solvents present in the final
formulation.
Environmental requirements
The table below describes the environmental requirements for Ignite and Ignite+
system.
Parameter Requirement
Allowed location Indoor use only
Ambient temperature 15°C to 35°C
Relative humidity 25% to 65%, non-condensing
Pressure 70 to 106 kPa
Altitude Up to 2000 m
9 Reference information
9.
1 Specifications
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 113